<6. SETTING AND CALIBRATION>
6-39
IM 11G02N01-01E
Settable range
<How to set/change the range>
ENT
( )
ENT
End of Set/Change the Range
ENT
ENT
( )
F6_8_4.ai
The measuring range can be arbitrarily selected in
the minimum and the maximum range specified at
the time of purchase. The range to be used can be
selected 1 or 2.
(1) Move the cursor to the item to be set by pressing
the
or the
key, and then press the
ENT
key.
(2) Move the cursor to the Ch (component) whose
setting is to be changed by pressing the
or
the
key, and then press the
ENT
key.
(3) Move the cursor to the item whose setting is to
be changed by pressing the
or the
key,
and then press the
ENT
key.
The value for range 1 and range 2 must fall within
the range from the MIN and the MAX range
(including the MIN and the MAX range), and
at the same time range 1 must be smaller than
range 2.
The number of ranges is 1 or 2.
(4) Press the
or the
key to change the
value. Press the
key to select the next
digit. In a state where the decimal point is
highlighted, press the
or the
key, and
the decimal point position can be changed.
(5) When necessary change is made, press the
ENT
key.