<6. SETTING AND CALIBRATION>
6-25
IM 11G02N01-01E
Description of setting items
6.6 Peak alarm setting
When the peak number of times CO concentration exceeds the upper limit value during measurement
exceeds the set number, an alarm is provided.
The peak alarm and this setting screen appear only when an option is added.
MODE
ENT
( )
ENT
( )
End of Peak Alarm Setting
ENT
( )
F6_6.ai
(1) Press the
MODE
key in the Measurement
mode, and the User mode appears.
(2) Point the cursor to “Setting of Peak
Alarm” by pressing the
or
key.
Press the
ENT
key.
(3) In the “Peak Alarm Setting” item selection
screen that appears, point the cursor
to any item you want to set by pressing
the
or
key. Press the
ENT
key.
(4) Then, enter numeric values and perform
the setting.
Entering the numeric values or setting the
items should be carried out by using the
or
key.
After setting, press the
ENT
key, and the set
values are saved.
• Peak Alarm : ON/OFF of peak alarm
• Alarm Value : If measuring value exceeds the
set alarm value, a peak counter
counts 1 time.
• Alarm Count : When a peak in excess of the
setting time occurs, a peak count
alarm output is provided.
• Hysteresis : To prevent possible chattering
when the measuring value may
exceed the set peak concentration
by only 1 time, the peak count has
an allowance in the hysteresis width.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...