<9. SPECIFICATIONS>
9-18
IM 11G02N01-01E
NO
2
/NO Converter
Part number:
K9350LE (Non-CE conformity), K9350LF (CE conformity)
Mounting:
Indoor surface mounting
Target Gases:
General boiler exhaust gas, atmosphere
Catalyst:
Amount; 2 cm
3
Replacement cycle;
Approx. 8 months (at flow rate of 0.5 L/min with 5%O
2
, 10 ppm NO)
Temperature setpoint; 220 ±5ºC (Sensing tip: K thermocouple)
Wetted materials:
Ceramic, Viton, glass filter, SUS316
Conversion efficiency: 90% or higher, conforms to JIS
Gas Flow Rate:
0.5 L/min
Ambient Temperature: -5 to +45ºC
Power Supply:
100 VAC , 50/60 Hz (K9350LE), 100 to 240 VAC , 50/60 Hz (K9350LF)
Power Consumption:
Approx. 85 VA
Safety and EMC conforming standards:
Installation altitude:
2000 m or less
Pollution degree:
2 (Note)
Installation category:
II (Note)
Note:
Installation category, called over-voltage category, specifies impulse withstanding voltage.
Category II is for electrical equipment. Pollution degree indicates the degree of existence of solid,
liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.
Safety:
EN61010-1
EMC:
EN61326-1 Class A, (For use in industrial locations),
EN61326-2-3, EN61000-3-2, EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand
Note:
The product mounted in a steel enclosure conforms to the requirements of EMC directive.
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Pleaseuse this instrument in the industrial environment only.
Weight:
Approx. 1.1 kg (K9350LE), Approx. 1.2 kg (K9350LF)
Sample gas requirements: Dust/drain removed, gas temperature at 150ºC or less
One-year-Use Spare Parts
Item
Part No.
Qty
Catalyst for NO
2
/NO converter
K9350LP
2
Glass wool for NO
2
/NO converter
K9350LQ
2
Fitting for NO
2
/NO converter
K9350LV
2
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...