<9.SPECIFICATIONS>
9-9
IM 11G02N01-01E
Model
Suffix code
Option code
Description
IR400
Infrared gas analyzer 19-inch rack mounting type with slide rail
4th Component
2nd Range
(note 2)
B
C
D
W
E
F
G
H
J
K
L
M
P
Q
R
S
T
U
N
0 - 100 ppm
0 - 200 ppm
0 - 250 ppm
0 - 300 ppm
0 - 500 ppm
0 - 1000 ppm
0 - 2000 ppm
0 - 5000 ppm
0 - 1%
0 - 2%
0 - 3%
0 - 5%
0 - 10%
0 - 20%
0 - 40%
0 - 50%
0 - 70%
0 - 100%
Not Available
O
2
Analyzer
1st Range
(note 2)
1
2
3
N
0 - 5%
0 - 10%
0 - 25%
Not available
O
2
Analyzer
2nd Range
(note 2)
2
3
N
0 - 10%
0 - 25%
Not available
Output
-4
-1
4 - 20 mA DC, non-isolation
0 - 1 V DC, non-isolation
Piping
R
T
Rc 1/4
1/4 NPT
Indication, Power
Cable (note 7)
J
E
U
Japanese, Power Cable; rated voltage 125 V AC
English, Power Cable; rated voltage 125 V AC (UL)
English, Power Cable; rated voltage 250 V AC (CEE)
Option
O
2
correction and O
2
average (note 4)
Communication
Internal purge (note 5)
Relay board (note 6)
/K
/A
/C
/P
/R
With O
2
correction and O
2
average value
With peak count alarm (note 4)
RS-232C (note 9)
Analyzer internal purging
With dedicated cable
Footnotes:
1:
A signal from the external O
2
analyzer should be 0-1 V DC linear to full scale.
2:
Possible combinations of ranges are specified in separate tables.
3:
Only available for CO
2
measurement. Option code “/P,” Analyzer internal purging, must be specified.
4:
O
2
correction is available only for NO, CO, and SO
2
. Both average value output after O
2
correction and average O
2
value output
are provided at the same time.
A peak count alarm can be provided only for CO measurement.
5:
When internal purging is specified with 3- or 4-component analyzers, only one set of gas inlet/outlet can be used and thus NO
2
/
NO converter cannot be connected between two measurement sections.
6:
Should be specified when using a solenoid valve for automatic calibration.
7:
Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select appropriate code according to the
operating power supply voltage to be used in the field.
Suffix Code “E” is of the North American plug type and “U” of the European type.
8:
For NOx measurement, a NO
2
/NO converter (P/N K9350LE or K9350LF) should be purchased separately.
9:
Should be specified when using Modbus
™
communication.
10:
ZX8D style D is specified for CE marking.
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...