<6. SETTING AND CALIBRATION>
6-38
IM 11G02N01-01E
• Other parameter
F6_8gamen_2.ai
Password Set : Set the password used to move
from the parameter setting screen
to the maintenance mode.
Arbitrary 4-digit number can be
selected.
O
2
ref. Value: Set the oxygen concentration
reference value at the time of
oxygen correction calculation.
Settable in the range from 00 to
19% (default 4%).
Limit :
Set the oxygen concentration limit
at the time of oxygen correction
calculation. Settable in the range
from 01 to 20% (default 17%).
* Refer to the O
2
correction concentration
value in “5.3 Outline of display screen” for
oxygen correction calculation procedure.
Station No. :
Set the station No. for MODBUS
communication. Settable in the
range from 00 to 32.
Range setting : Moves to the screen on which
measuring range is changed.
Press the
or the
key to move the
cursor to the item whose setting is to be
changed.
The values for password, oxygen
correction, limit, and station No. are
highlighted.
Press the
or the
key to change the
value to desired one, and then press the
ENT
key.
Note: Pay attention not to forget the password.
Otherwise you cannot enter the maintenance mode.
Description of each setting screen
Содержание IR400 series
Страница 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Страница 7: ...Blank Page ...
Страница 11: ...Blank Page ...
Страница 33: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 83: ...Blank Page ...
Страница 93: ...Blank Page ...
Страница 97: ...Blank Page ...
Страница 117: ...Blank Page ...