2-12
IM 701450-51E
Sample Point
Select a percentage for judging the bus level (recessive/dominant).
18.8, 21.9, 25.0, 28.1, 31.3, 34.4, 37.5, 40.6, 43.8, 46.9, 50.0, 53.1, 56.3, 59.4, 62.5,
65.6, 68.8, 71.9, 75.0, 78.1, 81.3, 84.4, 87.5, or 90.6 [%]
In the trigger circuits of the instrument’s CAN Bus signal analysis function, the input CAN
Bus signal is sampled once per the internal clock, and the point of change from
recessive to dominant is detected. The detected point is taken as 0%, and the point one
bit time (the reciprocal of the specified bit rate) thereafter is taken as 100%, allowing
expression of the sample point as a percentage.
1 bit
1 bit
62.5[%]
18.8[%]
90.6[%]
Sample point
When the Sample Point Is Set to 62.5%
Input waveform
Waveform
sampled at
the internal clock
Message Format
You can select a format for the Identifier (ID) field of the Arbitration field within the data
frame from the following:
Std
Standard format
Ext
Extended format
Vdiff
You can select a connection method for the differential probe from below. In either case,
the logical value is: recessive = 1 and dominant = 0.
CAN_L–CAN_H
The recessive voltage level is greater than the dominant voltage level.
CAN_H–CAN_L
The dominant voltage level is greater than the recessive voltage level.
Note
•
For the further information on connecting the differential probe, see section 2.3,
“Connecting the Probe.”
•
The bit rate setting is connected with the setting in the Analyze Setup dialog box (see
page 2-20), and the CURSOR menu (see page 2-35).
•
The Vdiff setting is connected with the setting in the Analyze Setup dialog box (see page
2-20).
Pattern Format
You can select a format for the Identifier (ID) field and the Data field from the following:
Hex
Hexadecimal
Bin
Binary
2.4 Setting CAN Bus Signal Acquisition Conditions