3-24
IM 701450-51E
3.5 Analyzing/Searching Data
Setting the Analysis Conditions
5.
Press the Analyze Setup soft key. The Analyze Setup dialog box opens.
Setting the Detection Level, Hysteresis, and Slope of the Synchronization Clock
Signal (CH1)
6.
Use jog shuttle & SELECT to set the level for detecting the synchronization
clock (Level box).
7.
Use jog shuttle & SELECT to set the hysteresis (width of the level used to
detect the synchronization clock) (Hysteresis box).
8.
Use jog shuttle & SELECT to select the slope, rising or falling, for detecting the
synchronization clock (Polarity box).
Set the hysteresis (width of
the level used to detect the
synchronization clock).
Set the level for detecting the synchronization clock.
Select which slope edge, rising or falling, of
the synchronization clock is to be detected.
Setting the Level for Determining the Status (1 or 0) of the I/O Data
9.
Use jog shuttle & SELECT to select whether Data1 (CH2) is to be analyzed
(ON/OFF) (Data1 (CH2) box).
•
Data1 (CH2) will be analyzed when the mark to the left is highlighted.
•
When interleave mode is ON, you cannot set Data1 (CH2).
10.
Use jog shuttle & SELECT to set the level for determining whether the status of
Data1 (CH2) is 1 (Thr Upper box).
11.
Use jog shuttle & SELECT to set the level for determining whether the status of
Data1 (CH2) is 0 (Thr Lower box).
•
When the data exceeds the specified level (Thr Upper), it is determined to be 1.
•
When the data is below the specified level (Thr Lower), it is determined to be 0.
•
When the data is between the levels specified by Thr Upper and Thr Lower (including
the Thr Upper and Thr Lower values), it is determined to be indefinite data.
12.
Repeat steps 9 to 11 to set the level for determining the status of Data2 (CH3).
Select whether the data is to be analyzed
Set the level for determining
whether the data status is 1.
Set the level for determining
whether the data status is 0.