2-5
IM 701450-51E
CAN Bus Signal Analysis Function
2
2.4
Setting CAN Bus Signal Acquisition
Conditions
With the CAN bus signal analysis function, you can acquire CAN Bus signals using
specific frames and fields of the CAN Bus as trigger conditions.
Procedure
•
To exit the menu during operation, press
ESC
located
above the soft keys.
• In the procedural explanation below, the term
jog shuttle &
SELECT
refers to the operation of selecting/setting items
and entering values using the
jog shuttle
,
SELECT
and
RESET
keys. For details on the operation using the jog
shuttle, SELECT, and RESET, see sections 4.1 or 4.2 in the
DL7440/DL7480 User’s Manual.
•
For a description of the operation using a USB keyboard or
a USB mouse, see section 4.3 in the DL7440/DL7480
User’s Manual.
SEARCH
PRESET
ZOOM
MODE
POSITION
ENHANCED
SIMPLE
LOGIC
CH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
V/DIV
TIME/DIV
VERTICAL
TRIGGER
TRIG D
ACTION
DELAY
CLEAR TRACE
SNAPSHOT
HELP
HORIZONTAL
1.
Press
ENHANCED
. The ENHANCED menu appears.
2.
Press the
Type
soft key. The trigger type selection menu appears.
3.
Press the
CAN Bus
soft key.
Setting the Trigger Conditions of the CAN Bus Signal
4.
Press the
CAN Setup