2-6
IM 701450-51E
Setting the Bit Rate
5.
Use
jog shuttle & SELECT
to display the keyboard used to set the bit rate (Bit
Rate box).
• When Setting the Bit Rate from Predefined Values
6.
Turn the
jog shuttle
to move the cursor onto the key that corresponds to the
desired bit rate (10, 20, 33.333, 50, 62.5, 83.333, 95.238, 100, 125, 250, 500, or
1000 [kbps]).
7.
Press
SELECT
. The selected bit rate is reflected on the keyboard.
Proceed to step 10.
• When Entering the Bit Rate Using Numeric Keys
6.
Press
RESET
to clear the current bit rate.
7.
Turn the
jog shuttle
to move the cursor to the highest digit of the bit rate you
wish to enter.
8.
Press
SELECT
.
9.
Repeat steps 7 and 8 to input the bit rate in the range of 10.000 to 1000.000
[kbps]. Proceed to step 10.
10.
Turn the
jog shuttle
to select ENT on the keyboard and press
SELECT
. The bit
rate is entered, and the keyboard disappears.
You can also press the
ENT
soft key to enter the bit rate and clear the keyboard.
Keyboard used to
set the bit rate
Note
•
If you input an arbitrary value using the numeric keys, the value that is actually entered
may differ from the value you specified. This is because, the bit rate resolution on the
instrument is set to the bit time (reciprocal of the bit rate) resolution of 0.5 [
µ
s], and the
instrument rounds the entered bit rate to the bit rate closest to the bit time conversion (see
page 2-11).
•
If you press RESET before the bit rate is entered, the bit rate you input is cleared.
•
Other operations on the keyboard are the same as the operations on the keyboard for
entering file names and comments. See page 4-4 in the
DL7440/DL7480 User’s Manual
(IM701450-01E)
.
2.4 Setting CAN Bus Signal Acquisition Conditions