C Bus Signal Analysis Function
1
To activate triggers on the start condition or address pattern/data pattern, proceed to
step 5 below. To activate triggers when an Acknowledge bit is not detected (Non-Ack
trigger), proceed to step 16 on the next page.
Activating Triggers on the;Start Condition or Address Pattern/Data Pattern
5.
Use jog shuttle & SELECT to select Adrs&Data (I
2
C Trigger Type box).
• Setting the Notation System of the Address Pattern or Data Pattern
6.
Use jog shuttle & SELECT to set the notation system used to set the pattern to
Hex or Bin (Pattern Format box). The format used to set the Address Pattern box,
Data1 Pattern box, and Data2 Pattern box are set to the specified notation system.
• Selecting the Items Used as Trigger Conditions
7.
Use job shuttle & SELECT to select whether Start Condition, Address Pattern,
Data1 Pattern, Byte Count, and Data2 Pattern are used as trigger conditions. If
you select Address Patten, proceed to step 8. If you select Data1 Pattern,
proceed to step 9. If you wish to set Byte Count, proceed to step 11. If you
select Data2 Pattern, proceed to step 12.
•
Highlighting of the mark to the left of each item indicates that it is used as a trigger
condition.
•
Start Condition and Byte Count items are always used as trigger conditions (the mark
to the left is always highlighted).
•
If you do not wish to use byte count as a trigger condition, set Byte Count to 0 (see
step 11).
• When Using Address Pattern as a Trigger Condition
8.
Use jog shuttle & SELECT to set the address pattern in hexadecimal or binary
(Address Pattern box). When determination is not to be performed, select X.
When using binary (BIN) notation, set the LSB using “R” (Read), “W” (Write), and “X.”
• When Using Data1 Pattern as a Trigger Condition
9.
Use jog shuttle & SELECT to set the determination pattern of Data1 in
hexadecimal or binary (Data1 Pattern box). When determination is not to be
performed, select X.
10.
Use jog shuttle & SELECT to set the Condition of the Data1 Pattern to True or
False (Condition box).
• When Setting the Byte Count
11.
Use jog shuttle & SELECT to set the byte count in the range of 0 to 9999 bytes
(Byte Count box).
•
If you do not wish to use byte count as a trigger condition, set the value to 0.
•
You can also set the byte count using the soft key that appears at the bottom of the
ENHANCED menu (a menu that appears when you carry out steps 1 to 3 on the
previous page). This byte count soft key menu appears only when I
2
C Trigger Type is
set to Adrs&Data.
Select Adrs&Data.
Select the setting format
of the address pattern
and data pattern.
The condition of activating
the trigger when a start
condition is detected is
always enabled.
To activate triggers on the
address pattern, select this
check box and set the
pattern.
To activate triggers on the
Data1 pattern, select this
check box and set the
pattern and conditions.
Set the byte count in the
range of 0 to 9999.
When Pattern Format is set to Bin
Set the LSB using R (Read), W (Write), or X.
1.4 Setting the I
2
C Bus Signal Acquisition Conditions