3-26
IM 701450-51E
Setting the Reference Point of the Analysis
19.
Use jog shuttle & SELECT to set the reference point to Trigger Position or
Manual (Trigger Position or Manual button).
The reference point is set to the one whose mark to the left is highlighted (Trigger
Position or Manual).
20.
If the reference point is set to Manual, use jog shuttle & SELECT to set the
reference point (Manual box on the right).
When setting the reference point using Manual, you can set the reference point while
checking the displayed waveform by enabling translucent display. For a description of
the translucent display, see section 8.10 in the
DL7440/DL7480 User’s Manual
(IM701450-01E).
Selecting the Read Direction of the I/O Data Bits
21.
Use jog shuttle & SELECT to set the read direction of the I/O data bits to MSB
First or LSB First (Bit Order box).
If the reference point is set to Manual,
set the reference point.
Set the reference point to Trigger Position or Manual.
Set the read direction of the I/O data bits
to MSB First or LSB First.
22.
Press ESC. The Analyze Setup dialog box closes.
Executing/Aborting the Analysis
23.
Press the Analyze Exec soft key. The analysis of the I/O data is executed. The
words Analyze Exec change to Analyze Abort.
To abort the analysis of the I/O data, press the Analyze Abort soft key. The
analysis of I/O data is aborted, and the words Analyze Abort change to Analyze
Exec.
If indefinite data exists in the analysis data, “
∗
” is attached to the corresponding
analysis data.
Data2 value
(hexadecimal)
Analysis result list display
CS signal status or the name of the CS signal whose priority is high
Analysis
number
Data1 value
(hexadecimal)
Displaying the Analysis Data
Scrolling the List
24.
Press the List soft key.
25.
Turn the jog shuttle to scroll the list left and right.
Up to 15 analysis data points are displayed at once in the order of occurrence. By
scrolling the list left and right using the jog shuttle, analysis data beyond 15 data points
can be displayed.
3.5 Analyzing/Searching Data