1-29
IM 701450-51E
I
2
C Bus Signal Analysis Function
1
Selecting the Search Type
25.
Use jog shuttle & SELECT to set the search type to Byte Pattern or Indefinite
State (Type box).
If you select Indefinite State, proceed to step 31.
26.
If you set the search type to Byte Pattern, use the jog shuttle & SELECT to set
the determination pattern to address pattern or data pattern.
•
Highlighting of the mark to the left of each item indicates that it is used as a
determination pattern.
•
You can select the address pattern and data pattern simultaneously.
Setting the Address Pattern
27.
Use the job shuttle & SELECT set the address pattern in hexadecimal (HEX)
or binary (BIN) format (HEX box or BIN box). When determination is not to be
performed, select X.
•
The HEX box and BIN box settings are synchronized.
•
When using binary (BIN) notation, set the LSB of the address pattern using “R” (Read),
“W” (Write), and “X.”
28.
Use jog shuttle & SELECT to set Ack to X, 1, or 0 (Ack box).
Setting the Data Pattern
29.
Use the job shuttle & SELECT set the data pattern in hexadecimal (HEX) or
binary (BIN) format (HEX box or BIN box). When determination is not to be
performed, select X.
The HEX box and BIN box settings are synchronized.
30.
Use jog shuttle & SELECT to set Ack to X, 1, or 0.
Set the search type to Byte Pattern or Indefinite State.
Set Ack to X, 1, or 0.
Set the address pattern in
hexadecimal or binary.
Set the data pattern in
hexadecimal or binary.
Set Ack to X, 1, or 0.
31.
Press ESC. The Search Setup dialog box closes.
1.5 Analyzing/Searching Data