3-46
IM 701450-51E
:SEARch:SPI:ANALyze:SETup:CS?
Function
Queries all settings related the chip select
signal (CS) in the analysis.
Syntax
:SEARch:SPI:ANALyze:SETup:CS?
Example
:SEARCH:SPI:ANALYZE:SETUP:CS? ->
:SEARCH:SPI:ANALYZE:SETUP:CS:CHANNEL4:
MODE 0;LEVEL 0.0E+00;:SEARCH:SPI:
ANALYZE:SETUP:CS:CHANNEL5:MODE 0;
LEVEL 0.0E+00;:SEARCH:SPI:ANALYZE:
SETUP:CS:CHANNEL6:MODE 0;
LEVEL 0.0E+00;:SEARCH:SPI:ANALYZE:
SETUP:CS:CHANNEL7:MODE 0;
LEVEL 0.0E+00;:SEARCH:SPI:ANALYZE:
SETUP:CS:CHANNEL8:MODE 0;
LEVEL 0.0E+00;:SEARCH:SPI:ANALYZE:
SETUP:CS:LOGIC:A0 0;A1 0;A2 0;A3 0;
A4 0;A5 0;A6 0;A7 0;:SEARCH:SPI:
ANALYZE:SETUP:CS:ESTATE NONE
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel<x>?
Function
Queries all settings related to each channel of
the chip select signal (CS) in the analysis.
Syntax
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel<x>?
<x>=4 to 8 (4 on the DL7440)
Example
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4?-> :SEARCH:SPI:ANALYZE:
SETUP:CS:CHANNEL4:MODE 1;
LEVEL 1.000E+00
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel<x>:LEVel
Function
Sets the level of each channel of the chip select
signal (CS) in the analysis or queries the
current setting.
Syntax
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel<x>:LEVel {<Voltage>}
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:LEVel?
<x>=4 to 8 (4 on the DL7440)
<Voltage>=8 divisions within the screen (0.01
division steps).
Example
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:LEVEL 1V
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:LEVEL? -> :SEARCH:SPI:
ANALYZE:SETUP:CS:CHANNEL4:
LEVEL 1.000E+00
Description
When a current probe is used, this command
sets or queries the <Current> value.
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel<x>:MODE
Function
Sets whether to handle the channel as a chip
select signal (ON/OFF) in the analysis or
queries the current setting.
Syntax
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:MODE {<Boolean>}
:SEARch:SPI:ANALyze:SETup:CS:
CHANnel4:MODE?
<x>=4 to 8 (4 on the DL7440)
Example
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:MODE ON
:SEARCH:SPI:ANALYZE:SETUP:CS:
CHANNEL4:MODE? -> :SEARCH:SPI:
ANALYZE:SETUP:CS:CHANNEL4:MODE 1
:SEARch:SPI:ANALyze:SETup:CS:LOGic?
Function
Queries all settings related to the logic input of
the chip select signal (CS) in the analysis.
Syntax
:SEARch:SPI:ANALyze:SETup:CS:LOGic?
Example
:SEARCH:SPI:ANALYZE:SETUP:CS:LOGIC?
-> :SEARCH:SPI:ANALYZE:SETUP:CS:
LOGIC:A0 0;A1 0;A2 0;A3 0;A4 0;
A5 0;A6 0;A7 0
Description
An error occurs if the logic input (optional) is not
installed.
:SEARch:SPI:ANALyze:SETup:CS:LOGic:
{A0|A1|A2|A3|A4|A5|A6|A7}
Function
Sets whether to handle each bit of the logic
input as a chip select signal (ON/OFF) in the
analysis or queries the current setting.
Syntax
:SEARch:SPI:ANALyze:SETup:CS:LOGic:
{A0|A1|A2|A3|A4|A5|A6|A7}
{<Boolean>}
:SEARch:SPI:ANALyze:SETup:CS:LOGic:
{A0|A1|A2|A3|A4|A5|A6|A7}?
Example
:SEARCH:SPI:ANALYZE:SETUP:CS:LOGIC:
A0 ON
:SEARCH:SPI:ANALYZE:SETUP:CS:LOGIC:
A0? -> :SEARCH:SPI:ANALYZE:SETUP:
CS:LOGIC:A0 1