6-11
IM AQ6370C-17EN
Status Registers
1
2
3
4
5
6
7
8
App
Index
Questionable Condition Register
Read
This register can be read by the :STATus:QUEStionable:CONDition? query command. Its
contents will not be cleared even when read.
Write
The register sets or resets a bit corresponding to a change in the status of the instrument
only when that change occurs. It cannot be written to.
Clear
The register cannot be cleared.
Questionable Event Register
Read
This register can be read by the :STATus:QUEStionable[:EVENt?] query command.
Its contents will be cleared when read.
Write
Contents of the register can be cleared. The register can be cleared but not written to.
Clear
This register will be cleared under any of the following conditions.
• A read using the :STATus:QUEStionable[:EVENt?] query command
• Initialization by the :STATus:PRESet command
• Common *CLS command
• Power ON
Questionable Event Enable Register
Read
This register can be read by the :STATus:QUEStionable:ENABle? query command.
Write
The register can be written to by the :STATus:QUEStionable:ENABle command.
Clear
This register will be cleared under any of the following conditions.
• Data “0” is set using the :STATus:QUEStionable:ENABle command.
• Power ON
The register cannot be cleared in the following cases.
• Receipt of the *RST command
• Receipt of the *CLS command
• Device clear (DCL, SDC)
6.5 Questionable Status Register