:CALCulate:AMARker[1|2|3|4]:FUNCtion
:PDENsity|:NOISe:BWIDth|:BANDwidth
Function
Sets or queries the normalization bandwidth of
the specified power spectral density marker.
Syntax
:CALCulate:AMARker:FUNCtion:PDENsit
y|:NOISe:BWIDth|:BANDwidth<wsp><NRf
>[m]
:CALCulate:AMARker:FUNCtion:PDENsit
y|:NOISe:BWIDth|:BANDwidth}?
<NRf> = Normalization bandwidth [m]
Example
:CALCULATE:AMARKER:FUNCTION:PDENSIT
Y:BWIDTH 0.1nm
:CALCULATE:AMARKER:FUNCTION:PDENSIT
Y:BWIDTH -> +1.00000000E-010
Description • This command applies to advanced markers
1 to 4. The command operates in the same
manner regardless of which advanced marker
is specified.
• An execution error will occur if the specified
advanced marker has not been assigned or is
not a power spectral density marker.
• A query error will occur if the specified
advanced marker has not been assigned or is
not a power spectral density marker.
• This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
:CALCulate:AMARker[1|2|3|4]:FUNCtion
:PDENsity|:NOISe:RESult?
Function
Sets or queries the power spectral density value
of the specified power spectral density marker.
Syntax
:CALCulate:AMARker[1|2|3|4]:FUNCtio
n:PDENsity|:NOISe:RESult?
[1|2|3|4]: Advanced marker number
Example
:CALCULATE:AMARKER:FUNCTION:PDENSIT
Y:RESULT? -> -1.00001
Description • The unit of the returned marker level depends
on the Y-axis unit of the assigned marker
trace.
• A query error will occur if the specified
advanced marker has not been assigned or is
not a power spectral density marker.
• If the advanced marker number is not
specified, advanced marker 1 will be used.
• This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
:CALCulate:AMARker[1|2|3|4]:FUNCtion
:PDENsity|:NOISe[:STATe]
Function
Sets the specified advanced marker to a power
spectral density marker.
Also queries the status of the specified power
spectral density marker.
Syntax
:CALCulate:AMARker[1|2|3|4]:FUNCtion:P
DENsity|:NOISe[:STATe]<wsp>OFF|ON|0|1
:CALCulate:AMARker[1|2|3|4]:FUNCtio
n:PDENsity|:NOISe[:STATe]?
[1|2|3|4]: Advanced marker number
Response 0 = Off, 1 = On
Example
:CALCULATE:AMARKER:FUNCTION:PDENSI
TY ON
:CALCULATE:AMARKER:FUNCTION:PDENSI
TY? -> 1
Description • If a power spectral density marker is assigned,
moving markers (:CALCulate:AMARker[1|2|3|4]
[:STATe]) and integrated power markers (:CAL
Culate:AMARker[1|2|3|4]:FUNCtion:INTegral[:
STATe]) will be set to off.
• If this command with the parameter set to
OFF is specified on an advanced marker that
has been assigned to power spectral density
marker, the advanced marker will change to a
moving marker.
If the advanced marker has not been
assigned, using this command with the
parameter set to OFF will leave the advanced
marker unassigned.
• If the advanced marker number is not
specified, advanced marker 1 will be used.
• This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
:CALCulate:AMARker[1|2|3|4]:FUNCtion
:PRESet
Function
Changes the specified advanced marker to a
moving marker.
Syntax
:CALCulate:AMARker[1|2|3|4]:FUNCtio
n:PRESet
[1|2|3|4]: Advanced marker number
Example
:CALCULATE:AMARKER:FUNCTION:PRESET
Description • If an advanced marker is assigned to a power
spectral density marker or integrated power
marker, the marker can be changed directly to
a moving marker.
This does not change the marker position.
If the advanced marker has not been
assigned, using this command will leave the
advanced marker unassigned.
• If the advanced marker number is not
specified, advanced marker 1 will be used.
• This is a sequential command.
• This command can be used on the AQ6370C/
AQ6370D/AQ6373B/AQ6375B.
7.6 Instrument-Specific Commands