AQ6317 Status Byte
The status byte of AQ6317-compatible mode operates like the status byte in the AQ6317.
Refer to the manuals for the AQ6317 series for the details of GP-IB.
Bit
Function and Setting Condition
Clear Timing
Bit 7
0
Bit 6
Send an SRQ signal.
• Upon execution of serial polling
• Upon receipt of DCL or SDC
Bit 5
When receiving data exceeding the receive • Upon execution of serial polling
buffer capacity of 512 byte “1” is set.
• Upon receipt of DCL or SDC
• At a start of measurement
Bit 4
0
Bit 3
When a command data error occurs, set “1”. • Upon receipt of DCL or SDC
• Upon execution of serial polling
• At a start of measurement
Bit 2
Warning error (including errors upon
• When the warning error display
execution of a Program) occurs, set “1”.
disappears
For the contents of the warning its number • Upon execution of serial polling
can be output
• Upon receipt of DCL or SDC
• At a start of measurement
Bit 1
When the execution of a copy or program
• Upon execution of serial polling
terminates, set “1”.
• Upon receipt of DCL or SDC
• At a start of measurement
Bit 0
After sweep finishes, “1” is set.
• Upon execution of serial polling
• Upon receipt of DCL or SDC
• At a start of measurement