2-4
IM AQ6370C-17EN
2.3 GP-IB Interface Specifications
GP-IB Interface Specifications
Electromechanical specifications: Conforms to IEEE std. 488-1978
Functional specifications:
See table below
Protocols:
Conforms to IEEE std. 488.2-1992
Encoding:
ISO (ASCII)
Mode:
Addressable mode
Address setting:
Addresses 0 to 30 can be set in the GP-IB setting
screen in the SYSTEM menu.
Remote mode cancel:
Press LOCAL to cancel Remote mode. Note that
this is disabled when under Local Lockout by the
controller.
Functional Specifications
Function
Subset Description
Source handshake
SH1
All capabilities of send handshake
Acceptor handshake
AH1
All capabilities of receive handshake
Talker
T6
Basic talker function, serial polling, and talker
cancel function through MLA (my listen address).
Talker only not provided.
Listener
L4
Basic listener function, serial polling, and listener
cancel function through MLA (my listen address).
Listener only not provided.
Service request
SR1
All service request functions
Remote local
RL1
All Remote/Local functions
Parallel port
PP0
Parallel polling function not provided
Device clear
DC1
All device clear functions
Output buffer clear
Input buffer clear (clearing of an unexecuted
commands)
Error buffer clear
STB, ESR clear
Device trigger
DT0
Device trigger function
Controller
C0
Controller function not provided
Electrical characteristics
E1
Open collector