App-17
IM 765501-01E
Appendix
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
17
16
Index
App
App
Procedure
Preparation
Set the GS610 as follows:
Setup Example of the GS610
SOURCE
Output function
Current output
Limiter value
3 V
Output mode
Pulse
Output range
AUTO
OUTPUT CONTROL
Trigger mode
Internal
MEASURE
Measurement function
Voltage measurement
Integration time
250
µ
s
Measurement range
AUTO
AUTO V/I
ON
(auto switching of the measurement mode)
Store count
65535 points
TIME
Source delay
1
µ
s
Measurement delay
1
µ
s
Repeat period (of the internal trigger) 40 ms
Pulse width
5 ms
Source and Measurement
1.
The GS610 outputs (sinks) the pulse current to the secondary battery and
measures the voltage.
The source values and measured results are saved to a single file in the internal
memory of the GS610 (GS610RAM).
2.
Transfer (Copy) the measurement result file from the GS610RAM to the PC via
the USB.
Appendix 2 Application Examples