3-8
IM 765501-01E
Connecting the Power Cord
1.
Check that the power switch is OFF.
2.
Connect the power cord plug to the power connector on the rear panel. (Use
the power cord that comes with the package.)
3.
Connect the plug on the other end of the power cord to the outlet that meets the
conditions below. The AC outlet must be of a three-prong type with a protective
earth ground terminal.
Item
Specifications
Rated supply voltage
100 to 120 VAC, 220 to 240 VAC
Permitted supply voltage range
90 to 132 VAC, 198 to 264 VAC
Rated power supply frequency
50/60 Hz
Permitted supply voltage frequency range
48 to 63 Hz
Maximum power consumption
Approx. 200 VA
* The instrument can use a 100-V or a 200-V system for the power supply. Check that the
voltage supplied to the instrument is less than or equal to the maximum rated voltage of the
provided power cord (see page iv) before using it.
3-prong outlet
Power cord
(standard accessory)
USB
100 - 120V
220 - 240V A
C
200V
A MAX
50
60Hz
EXT I
O
3.3 Connecting to the Power Supply