18-2
IM 765501-01E
Response Time (Typical)
Voltage Source
100
µ
s: Time to reach within
±
0.1% of the setting
1
(at 20 V range setting with the source and limiter settings at maximum values and with a
pure resistive load with a load factor of 25%)
1 Measure delay needed for the measured voltage to reach
±
0.1% in pulse source
mode with the pulse base value set to zero.
1 Integration time: 250
µ
s
Current Source
400
µ
s: Time to reach within
±
1% of the setting
2
(at 20 A range setting with the source and limiter settings at maximum values and with a
pure resistive load)
2 Measure delay needed for the measured current to reach
±
1% in pulse source mode
with the pulse base value set to zero.
2 Integration time: 250
µ
s
Output Noise (Typical)
8 mVp-p (DC to 20 MHz)
(with 2 V source range setting and at 1 A limiter range)
18.1 Source Section