11-4
IM 755601-01E
Error in System Operation (901 to 905)
Code Message
Probable Cause/Corrective Action
Page
900
ROM checksum error
Servicing is required
–
901
Backup lithium battery low
Servicing is required for battery replacement. –
902
Setup param. checksum error (other than comm.) Servicing is required
–
903
Setup param. checksum error (comm.)
Servicing is required
–
905
EEPROM checksum error
Servicing is required
–
If an error code in the range 902 to 905 is displayed at power-up, the error remains on the screen until a key is pressed.
For system errors (900 to 905), the "
:STATus:ERRor?
" query cannot be used to inquire error codes and error messages.
Error in System Operation (Communications, 912)
Code Message
Probable Cause/Corrective Action
Page
912
Fatal error in Communication-driver
Servicing is required
–
Warning (Communications, 5)
Code Message
Probable Cause/Corrective Action
Page
5
*OPC/? Exists in message
Place the
*OPC
or
*OPC?
command at the
–
end of the program message.
Miscellaneous (-350, -390)
Code Message
Probable Cause/Corrective Action
Page
-350 Queue overflow
This error only occurs for the
10-40
"
:STATus:ERRor?
" query.
The error code does not appear on the
instrument’s display.
-390 Overrun error (RS232 only)
Lower the baud rate.
–
This message sometimes appears if a
command is sent during the [measurement
time].
11.2 Messages and Corrective Actions