Chapter 4 Setting the Measurement Conditions
LOCAL key
GP-IB connector
4.1 Switching the Limit Mode
Keys
NORMAL
MEAS&
LOCK
SAMPLE
EXTERNAL
REFERENCE
MANUAL
FAST
HI
HI LIMIT
EXIT
SETUP/MEAS
CHECK
MISC
I/F
MEAS
TIME
STORE
DATA
RECALL
DATA
LOCAL
ENTER
LIMIT
MODE
TRIG
MODE
SHIFT
R/%
REF
LIMIT
TRIG
7
1
0
.
2
3
5
4
6
8
9
LO LIMIT
IN
LO
M
Ω
M
Ω
k
Ω
Ω
,%
k
Ω
Ω
Ω
%
%
HIGH
SPEED
STORE
DATA
RECALL
DATA
CHECK
REMOTE
Indicator is off during SETUP mode
Procedure
The items that are specified or selected are confirmed when the ENTER (
Ω
,%) key is
pressed. To exit from a menu in the middle of the operation, press the EXIT (R/%) key.
Switching to the SETUP mode
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the SETUP mode. If the “MEAS & LOCK”
indicator is off, this operation is not necessary.
Switching the limit mode
2.
4.
3.
(Deviation mode)
(Absolute mode)
ENTER
LIMIT
MODE
SHIFT
REF
LIMIT
8
Ω
,%
[Main display]
Explanation
There are two limit modes.
You can select whether to use a deviation (%) or an absolute value (R) for the
comparator function.
• Deviation (%) mode: The measured value is handled as a deviation from the
specified reference value. The comparator function is also
carried out in terms of the deviation.
• Absolute (R) mode:
The measured value is handled as an absolute value. The
comparator function is also carried out in terms of the absolute
value.
Precautions to be taken when switching the limit mode
The following parameters are initialized when the limit mode is switched.
HI level, LO level, and REF (reference, when the limit mode is set to %)