10-31
IM 755601-01E
Communication
Commands
10
:STORe:POINts?
Function
Queries the number of data points that are
stored.
Syntax
:STORe:POINts?
Example
:STORE:POINTS?
→
0
Response
<NR1> form
Description Outputs the number of measured data points
that are currently stored in the internal memory.
:STORe[:STATe]
Function
Sets the measured data store state (ON/OFF)
or queries the current setting.
Syntax
:STORe[:STATe] {<Boolean>}
:STORe:STATe?
Response
{1|0}
Example
:STORE:STATE ON
:STORE:STATE?
→
:STORE:STATE 1
Description • Executing the “
:STORe[:STATe] ON
”
command sets the instrument in a store start
ready state. The actual store operation is not
executed at this point. Measured data are
stored into the memory every time a trigger
occurs after the instrument is switched to the
measurement mode using the “
:MEASure ON
”
command.
• When the number of data points specified by the
“
:STORe:COUNt
” command is stored, the
“
:STORe[:STATe]
” is automatically turned OFF.
Executing the “
:STORe[:STATe] OFF
” command
will forcibly terminate the store operation.
• If “
:STORe[:STATe]
” is turned ON when there
are measured data already in the memory, all
of the stored data are cleared.
• If the “
:RECall[:STATe]
” is turned ON,
“
:STORe[:STATe]
” cannot be turned ON.
Attempting to do so will result in an error
(831).
10.2 Commands