12-4
IM 755601-01E
12.4 Contact Check Specifications
Item
Specifications
Check level setting range
1 to 30
Ω
(resolution: 1
Ω
)
Check time
2 ms
Execution timing
Select before or after the measurement
Test current
50 mA
When a contact check occurs
Display
: “–nC–”
Determination
: HI
Handler interface
: NO CONTACT, HI output
12.5 Comparator Specifications
Setting Range
Parameter
Absolute Setting Range
Deviation Setting Range
Percent Limit: 9.99
Percent Limit 99.9
HI
0.0000
Ω
to 1.2000
Ω
-9.999% to 9.999%
*3
-99.99% to 99.99%
*4
(for 755611)
*1
LO
0.0000
Ω
to 1.2000
Ω
-9.999% to 9.999%
*3
-99.99% to 99.99%
*4
(for 755611)
*2
Set so that LO is less than or equal to HI
*1, *2 0.000
Ω
to 1.200
Ω
for 755601
*3
-9.99% to 9.99% for 755601
*4
-99.9% to 99.9% for 755601
Comparison
The signal output behavior of INDEX, EOM, NO CONTACT, HI, IN and LO signals vary depending on the trigger mode as follows.
When the trigger mode is set to EXTERNAL
INDEX:
Output when a trigger occurs regardless of the measurement result.
EOM:
Output once when the measurement completes regardless of the measurement result.
NO CONTACT:
Output when the measurement result is a contact check error (nC) or when abnormalities are detected in
the test current (CF).
HI:
Output when the comparator result of the measured value is HI.
Output also when the measurement result is an overrange (oL), contact check error (nC), or when
abnormalities are detected in the test current.
IN:
Output when the comparator result of the measured value is IN.
LO:
Output when the comparator result of the measured value is LO.
When the trigger mode is set to MANUAL or INTERNAL
INDEX:
Not output (Stays at H).
EOM:
Not output (Stays at H).
NO CONTACT:
Output when the measurement result is a contact check error (nC) or when abnormalities are detected in
the test current (CF).
Output when the comparator result of the measured value is HI.
Output also when the measurement result is an overrange (oL), contact check error (nC), or when
abnormalities are detected in the test current.
IN:
Output when the comparator result of the measured value is IN.
LO:
Output when the comparator result of the measured value is LO.