10-20
IM 755601-01E
10.2.7
LIMit Group
The commands in the LIMit Group deal with the limit mode and limit values. These commands can be used to make the same
settings and inquiries as the “PLmt” menu accessible through the LIMIT MODE, REF, LIMIT, and MISC key on the front panel.
:LIMit
:
MODE
<Space>
PCNT
OHM
?
PCNT
:
REFerence
<Space>
?
PLIMit
<Space>
<NRf>
?
DATA
<Space>
<NRf>
,
<NRf>
?
?
OHM
:
DATA
<Space>
,
?
?
?
<Resistance>
<Resistance>
<Resistance>
;
;
:LIMit?
Function
Queries all settings related to the limit mode
and limit values.
Syntax
:LIMit?
Example
• When the limit mode is set to deviation mode (%)
:LIMIT?
→
:LIMIT:MODE PCNT;PCNT:
REFERENCE 1.0000E+05;PLIMIT 9.99;DATA
5.00,—5.00
• When the limit mode is set to absolute mode (R)
:LIMIT?
→
:LIMIT:MODE OHM;OHM:DATA
1.0500E+00,0.9500E+00
:LIMit[:MODE]
Function
Sets the limit mode or queries the current setting.
Syntax
:LIMit[:MODE] {PCNT|OHM}
:LIMit:MODE?
PCNT = Deviation mode (%)
OHM = Absolute mode (R)
Response
{PCNT|OHM}
Example
:LIMIT:MODE PCNT
:LIMIT:MODE?
→
:LIMIT:MODE PCNT
Description When the “
:LIMit[:MODE]
” setting is changed,
the following settings are also automatically
changed as follows.
When changed from OHM to PCNT
Display unit “
:DISPlay[:MODE]
” Deviation (%)
Limit values “
:LIMit:PCNT[:DATA]
”
HI=0%,
LO=0%
Ref. val “
:LIMit:PCNT:REFerence
”
100.00 k
Ω
When changed from PCNT to OHM
Display unit “
:DISPlay[:MODE]
” Absolute(R)
Limit values “
:LIMit:OHM[:DATA]
”
HI= 0
Ω
,
LO= 0
Ω
:LIMit:OHM?
Function
Queries all settings when the limit mode is set
to absolute mode (R).
Syntax
:LIMit:OHM?
Example
:LIMIT:OHM?
→
:LIMIT:OHM:DATA 1.0500E+00,
0.9500E+00
:LIMit:OHM[:DATA]
Function
Sets the limit values (absolute value (R)) or
queries the current setting.
Syntax
:LIMit:OHM[:DATA] {<Resistance (HI)>,
<Resistance (LO)>}
:LIMit:OHM:DATA?
<Resistance> = 0.000
Ω
to 1.200
Ω
Response
<NR3>, <NR3> (Output in the order HI and
LO)
Example
:LIMIT:OHM:DATA 1.05OHM,0.95OHM
:LIMIT:OHM:DATA?
→
:LIMIT:OHM:DATA
1.0500E+00, 0.9500E+00
Description • The HI value must be greater than the LO
value. Otherwise, an error (815) occurs.
• When the “
:LIMit[:MODE]
” is set to PCNT,
the “
:LIMit:OHM[:DATA]
” setting cannot be
changed or queried. Doing so will result in
an error (812).
10.2 Commands