10-18
IM 755601-01E
:COMMunicate:WAIT?
Function
Creates the response that is returned when the
specified event occurs.
Syntax
:COMMunicate:WAIT? <Register>
<Register> = 0 to 65535 (Extended event
register, see page 10-39)
Response
1 (Fixed)
Example
:COMMUNICATE:WAIT? 65535
→
1
:DISPlay[:MODE]
Function
Sets the displayed unit (deviation (%) or
absolute (R)) of the measured value or queries
the current setting.
Syntax
:DISPlay[:MODE] {PCNT|OHM}
:DISPlay:MODE?
PCNT = Deviation (%) display
OHM = Absolute (R) display
Response
{PCNT|OHM}
Example
:DISPLAY:MODE PCNT
:DISPLAY:MODE?
→
:DISPLAY:MODE PCNT
Description When the “
:LIMit[:MODE]
” is set to OHM, this
command cannot be used to specify PCNT.
Doing so will result in an error (814).
10.2.4
DISPlay Group
The commands in the DISPlay Group deal with the measurement display.These commands can be used to make the
same settings and inquiries as when the R/% key on the front panel is pressed.
:DISPlay
:
MODE
<Space>
PCNT
OHM
?
?
:DISPlay?
Function
Queries all settings related to the measurement
display.
Syntax
:DISPlay?
Example
:DISPLAY?
→
:DISPLAY:MODE PCNT
10.2.5
HANDler Group
The commands in the HANDler Group deal with the handler interface. These commands can be used to make the
same settings and inquiries as the “Eom” menu accessible through the MISC key on the front panel.
:HANDler
:
EOM
<Space>
MAX
MIN
?
?
<Time>
:HANDler?
Function
Queries all settings related to the handler
interface output.
Syntax
:HANDler?
Example
:HANDLER?
→
:HANDLER:EOM 0.0100
:HANDler:EOM
Function
Sets the pulse width of the EOM signal or
queries the current setting.
Syntax
:HANDler:EOM {<Time>|MAX|MIN}
:HANDler:EOM?
<Time> = 0.1, 5, 10, 15(ms)
MAX = 15ms
MIN = 0.1ms
Response
{0.0001|0.0050|0.0100|0.0150} (<NR2>
form)
Example
:HANDLER:EOM 10MS
:HANDLER?Æ:HANDLER:EOM 0.0100
10.2 Commands