4-6
IM 755601-01E
When the limit mode is set to absolute (R)
Keys
NORMAL
MEAS&
LOCK
SAMPLE
EXTERNAL
REFERENCE
MANUAL
FAST
HI
HI LIMIT
EXIT
SETUP/MEAS
CHECK
MISC
I/F
MEAS
TIME
STORE
DATA
RECALL
DATA
LOCAL
ENTER
LIMIT
MODE
TRIG
MODE
SHIFT
R/%
REF
LIMIT
TRIG
7
1
0
.
2
3
5
4
6
8
9
LO LIMIT
IN
LO
M
Ω
M
Ω
k
Ω
Ω
,%
k
Ω
Ω
Ω
%
%
HIGH
SPEED
STORE
DATA
RECALL
DATA
CHECK
REMOTE
Set/Display HI limit value
Set/Display LO limit value
Indicates the comparator result
Procedure
The items that are specified or selected are confirmed when the ENTER (
Ω
,%) key is
pressed. When confirmed, the new reference is displayed.
To exit from a menu in the middle of the operation, press the EXIT (R/%) key. The
display returns to the top menu of the SETUP mode.
Switching to the SETUP mode
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the SETUP mode. If the “MEAS & LOCK”
indicator is off, this operation is not necessary.
Changing the limit values
4.
2.
3. Numerical keys
5. Numerical keys
6.
LIMIT
ENTER
Ω
,%
ENTER
Ω
,%
[HI LIMIT display]
[LO LIMIT display]
If the value entered using the numerical keys is not correct, an error (810) is displayed
for approximately one second, and the display returns to step 3.
Explanation
Setting range of limit values
The setting range applies to both HI and LO.
Model
Setting Range
Resolution
755601 0.000
Ω
to 1.200
Ω
1 m
Ω
755611 0.0000
Ω
to 1.2000
Ω
100
µ
Ω
However, LO
≤
HI
Note
If the values are set so that HI becomes less than LI, an error (815) occurs.
Comparator function
Same as the description given in “When the limit mode is set to deviation (%).” See
page 4-5.
4.3 Using the Comparator Function