4.5
What to do if... / printer
Integrated printer
does not print
Printer operating -
paper not being
printed
Integrated printer
prints automatically
Cause
Remedy
– Printer switched off
(LED flashing)
– Switch on the printer
(LED illuminated)
– No power pack connected
– Connect the power pack
– USB cable connected
– Disconnect the USB cable from the
meter
– Function "store automatically at
intervals" with long intervals is
switched on
– Switch off the function (see section
– No paper available
– Insert a roll of paper
Cause
Remedy
– Paper inserted the wrong way up – Turn the roll of paper around and
insert it with the other side upwards
Cause
Remedy
– The function "store automatically
at intervals" or "transmit data
automatically at intervals" is
switched on
– Switch off the functions