pH value
TruLab pH/ISE 1320
36
ba76161e01
08/2015
Continuing with
three- to five-point
calibration
17.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the setting of the nominal buffer value
appears.
18.
Set the nominal buffer value for the measured temperature with
<
><
>
.
19.
Accept the set calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
20.
Finish the calibration procedure as a two-point calibration with
<M>
.
The calibration record is displayed.
pH
4.035
24.8
°C
ConCal
2
12.12.2012
12:12
Buffer
21.
Thoroughly rinse the electrode with deionized water.
22.
Immerse the electrode in the next buffer solution.
23.
When measuring without temperature sensor:
Enter the temperature of the buffer with
<
><
>
.
24.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the setting of the nominal buffer value
appears.