TruLab pH/ISE 1320
pH value
ba76161e01
08/2015
39
Calibration record
(example)
Display
Calibration
record
Zero point
[mV]
Slope
[mV/pH]
+++
- 15 ... + 15
-60.5 ... -58.0
++
- 20 ... <-15
or
>+15 ... + 20
>-58.0 ... -57.0
+
- 25 ... <- 20
or
>+ 20 ... + 25
-61.0 ... <-60.5
or
>-57.0 ... -56.0
-
- 30 ... <- 25
or
->+25 ... + 30
-62.0 ... <-61.0
or
>-56.0 ... -50.0
Clean the electrode according to the elec-
trode operating manual
Error
Error
- 30 ... + 30
-62.0 ... -50.0
Error elimination (see section 13 W
pH/ISE 1320
Ser. no. 11292113
CALIBRATIONpH
12.12.2012 15:55
Ser. no. 10501234
AutoCal TEC
Buffer 1
4.00
Buffer 2
7.00
Buffer 3
10.00
Voltage 1
184,0 mV
Voltage 2
3,0 mV
Voltage 3
-177,0 mV
Temperature 1
24.0 °C
Temperature 2
24.0 °C
Temperature 3
24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV