pH value
TruLab pH/ISE 1320
30
ba76161e01
08/2015
6.1.2 Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
(NTC 30 or Pt 1000) integrated in the sensor.
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with temperature sensor
-
1 °C
Manual
1.
Measure the current temperature of the test sample.
2.
In the
<MENU/ENTER>
/pH/
Man. temperature
menu, set the tempera-
ture value with
<
><
>
.
Содержание YSI pH/ISE 1320P