TruLab pH/ISE 1320
Ion concentration
ba76161e01
08/2015
49
Continuing with two-
point calibration
7.
Thoroughly rinse the electrode with distilled water.
8.
Immerse the electrode in standard solution 1.
9.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
<F2>
/[ T
⇅
C ] to select the setting of the temperature.
Use
<
> <
>
to set the temperature.
Use
<F2>
/[ T
⇅
C ] to select the setting of the concentration.
10.
Set the concentration of the standard solution with
<
> <
>
and
press
<MENU/ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
11.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with
<MENU/ENTER>
.
The calibration display for the next standard solution appears.
ISE
0.0
24.8 °C
12.12.2012
12:12
0.01 mg/
mV
AR
ISE
0.0
24.8 °C
Ag
0.020
mg/l
2
12.12.2012
12:12
Standard
mV
12.
Thoroughly rinse the electrode with distilled water.
13.
Immerse the electrode in standard solution 2.