TruLab pH/ISE 1320
Ion concentration
ba76161e01
08/2015
65
Measuring
CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l
20.2 mV 25.0 °C
#2 0.100 mg/l
79.2 mV 25.0 °C
Slp.: 59.0 mV
++ (Ion type: Ag)
12.12.2012
12:12
Continue
USB output
5.
Start the measurement with
<F1>
Continue
An entry window appears.
6.
Prepare the sample.
7.
If necessary, add the ISA/TISAB solution to the sample.
8.
Thoroughly rinse the electrode with deionized water.
9.
Immerse the electrode in the sample that was supplemented with blank
value solution.
10.
Using
<
> <
>
and
<MENU/ENTER>
, enter the values for the vol-
ume of the sample (
Sample volume
), the volume of the ISA/TISAB
solution (
ISA/TISAB vol.
), the volume of the blank value solution (
BV
volume
) and the concentration of the blank value solution (
BV conc.
).
11.
Select
Continue
with
<
> <
>
and start the measurement with
<MENU/ENTER>
.
When the measurement is finished an entry window appears.
i
Immerse sensor in sample
Sample volume
100.0 ml
ISA/TISAB vol.
1.0 ml
BV volume
100.0 ml
BV conc.
1.000 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back
i
Add standard!
Std. volume
100.0 ml
Std. conc.
1.0 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back