Data storage
TruLab pH/ISE 1320
82
ba76161e01
08/2015
Energy saving mode
([Eco mode])
If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
saving mode switches off functions of the meter that are not required for the
automatic storage of measurement data (such as the display). By pressing any
key the energy saving mode is switched off again.
Terminating the
automatic storage
function
prematurely
Proceed as follows to switch off the automatic data storage function before the
adjusted storage duration has expired:
10.3 Measurement data storage
10.3.1 Editing the measurement data storage
The contents of the manual or automatic measurement data storage can be
shown on the display and output to the interface.
Each of the measurement data storages has a function to erase the entire
contents.
Editing the data
storage
The storage is edited in the menu,
Storage & config
/
Data storage
. To open
the
Storage & config
menu, press the
<MENU/ENTER__>
key in the measured
value display.
Open the manual or automatic storage directly with the
<RCL>
or
<RCL__>
key.
1.
Press the
<STO__>
key.
The following window appears.
2.
Using
<
><
>
, select
Yes
and confirm with
<MENU/ENTER>
.
The meter switches to the measured value display.
The automatic data storage function is terminated.
Stop automatic storage?
Yes
No
Warning
12.12.2012
12:12
Back
The settings are explained here using the manual data storage as
an example. The same settings and functions are available for the
automatic data storage.
Содержание YSI pH/ISE 1320P
Страница 6: ...Contents TruLab pH ISE 1320 6 ba76161e01 08 2015...
Страница 104: ...Contact Information TruLab pH ISE 1320 104 ba76161e01 08 2015...
Страница 105: ......