Ion concentration
TruLab pH/ISE 1320
58
ba76161e01
08/2015
10.
Using
<
> <
>
and
<MENU/ENTER>
, select the values for the vol-
ume of the sample (
Sample volume
) and the volume of the ISA/TISAB
solution (
ISA/TISAB vol.
).
11.
Select
Continue
with
<
> <
>
and start the measurement with
<MENU/ENTER>
.
When the measurement is finished an entry window appears.
12.
Add the standard (complexing agent or precipitating agent) to the sam-
ple.
13.
Using
<
> <
>
and
<MENU/ENTER>
, enter the values for the vol-
ume of the standard solution (
Std. volume
) and concentration of the
standard solution (
Std. conc.
) .
14.
If necessary, set the ion type for the ion in the standard (
Ion type
) with
<
> <
>
and
<MENU/ENTER>
.
On selection of the ion type that can be defined (ION):
Using
<
> <
>
and
<MENU/ENTER>
, set the valence (
Valency
) and
molar weight (
Molar mass
) for the ion in the standard solution.
15.
Select
Continue
with
<
> <
>
and start the measurement with
<MENU/ENTER>
.
The measurement result is displayed when the measurement is com-
pleted.
16.
If necessary, start measuring further samples with
<MENU/ENTER>
.
Repeat steps 6 - 13 for all samples.
17.
Terminate the measuring method with
<M>
.
A safety query appears.
i
Add standard!
Ion type
S
Std. volume
1.0 ml
Std. conc.
1.0 mg/l
Continue
Standard subtraction
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Standard subtraction
mg/l
USB output