Ion concentration
TruLab pH/ISE 1320
52
ba76161e01
08/2015
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Display
Calibration
record
Calibration evaluation
Magnitude of the slope [mV]
+++
Ideal
50.0 ... 70.0 *
or 25.0 ... 35.0 **
-
Poor
(measurement accuracy restricted)
Eliminate sources of error
(see section 13.2 ISE, page 91).
Recalibrate
30.0 ... 50.0 or 70.0 ... 90.0 *
or 15.0 ...25.0 or 35.0 ... 45.0 **
Error
Error
Insufficient
(measuring not possible)
Error elimination (see section 13.2
ISE, page 91)
< 30 or > 90.0 *
or < 15 or > 45 **
*
monovalent ions
** divalent ions