Alyza IQ PO4
What to do if ...
155
ba76201d03
05/2020
Implausible
measured values
Cause
Remedy
Erroneous measurement
Wait for another measurement
Calibration error
Repeat calibration
Check the calibration standards
Check the connection of the calibra-
tion standards
(for 1-point calibration the ChemBag
at connector S1 is always used.)
Refill the tubes for the ChemBags
with standard solutions
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Befüllen (einzeln)
Exchange calibration standards
Increase the time between cleaning
and calibration.
Measured values always too
high after cleaning
Activate the function
Konditionierung
(menu
Einstellungen Sensoren/Dif-
ferenzsensoren
/
Autom.Reinigung
).
After cleaning, a rinse is carried out for
each of the set number of steps and then
a rinse with the sample and the added
reagent.
Measuring cell of the photometer
unit dirty
Carry out a cleaning cycle
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Reinigen
Contact the service department
Interferences
e.g. in sample matrix
Activate the function
Untergr.-korr.
(opt)
.
PLEASE NOTE:
For the optimized background correc-
tion, an extra tube and an extra
ChemBag have to be installed (see
section 5 Maintenance and cleaning,
Select suitable measuring location
Photometer unit draws in air (e.g.
air bubbles in the sample tube)
Reduce the pump capacity of the fil-
tration pump (set to approx. 2 - 5 %).
Check the screw couplings of the
tubes for tightness.