Alyza IQ PO4
Maintenance and cleaning
103
ba76201d03
05/2020
11
Make sure that the connected ChemBags are safely suspended from
the supporting rod, and that they are not pressed or moved while the
locking device is open.
12
Make sure that the sample tubes (from the overflow vessel to the mea-
suring unit) are fixed in the recesses at the left-hand side of the measur-
ing unit.
13
Keep absorbent and lint free paper at hand so you can immediately
absorb any small chemicals remains.
14
Open the locking device of the MultiPort valve (MPV)
Beeps indicate that the locking device is open.
When the locking device is open, there is the risk of leaking chemicals.
15
Immediately place the absorbent, lint free paper onto the MultiPort
valve.
16
Move the liquid interface above the MultiPort valve a little bit downward
by hand so that it contacts the paper, and liquid remains are absorbed.
The locking device is now safely opened.
17
Make sure that the seals of the liquid interface remain within the inter-
face.
figure 5-2 Removing the liquids remains with absorbent, lint free paper
Содержание WTW Alyza IQ PO4
Страница 171: ......