Maintenance and cleaning
Alyza IQ PO4
118
ba76201d03
05/2020
figure 5-9 Connecting the ChemBag
1
Color coding of the tube
2
Color coding of the ChemBag
39
Carefully turn the ChemBag upside down.
The connector points downward.
40
If necessary, slightly knock on the connector of the ChemBag to remove
any small bubbles from the connector.
41
Suspend the ChemBag from the supporting rod.
42
Connect all selected ChemBags
– 1 cleaning solution (blue)
– 1 - 2 standard solutions (orange)
– 1-point calibration: Standard solution to connector S1
– 2-point calibration:
low standard solution to connector S1,
higher standard solution to connector S2
– 1 - 2 reagent solutions (green)
43
For each ChemBag, enter the expiry date and, if necessary, the filling
level (for used ChemBags) on the display.
44
Guide all tubes of the ChemBags over the tube fasteners above the
locking device and fix the tubes in the intervals of the tube fasteners.
The Alyza IQ has an extra counter for each ChemBag (see
section 1.3.3 ChemBags,
When delivered, the counters are set to 100 %.
1
2