What to do if ...
Alyza IQ PO4
152
ba76201d03
05/2020
6
What to do if ...
Display “----”
Information on the cause of the indication is in the log book and in
the Alyza menu / tab
Status
.
Cause
Remedy
No (valid) measured value
available
Start measurement and wait for the mea-
surement to be finished (5 ... 7 min)
Three erroneous automatic
calibrations in succession
Check the calibration standards
(e.g. expiry date)
Check the connection of the calibration
standards
(for 1-point calibration the ChemBag at
connector 'S1' is always used.)
Refill the tubes for the ChemBags with
standard solutions
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Befüllen (einzeln)
Carry out a cleaning cycle
Alyza menu / tab
Wartung
/
Manuelle
Funktionen
/
Reinigen
Exchange calibration standards
Contact the service department
Status
FEHLER
The Alyza IQ is stopped
General proceeding
View the details of the error
e.g. in the Alyza menu / tab
Status
(dis-
play details with
<OK>
)
or
View the log book message
Reset the error:
(Alyza menu, tab
Wartung
/
Manuelle
Funktionen
/
Fehler zurücksetzen
)
Remedy the error
Alyza IQ STARTEN
If the
Status
FEHLER
is still there:
Contact the service department