NOTE: The centrifugal pump
alerts listed on this page indicate
Low
-
priority (static yellow) alarms
which will not cause pump stop-
page, but need to be evaluated/
addressed.
6.2.2 CardioHelp
XVIVO
(Low
-
priority)
Alert
Definition
Action Needed
Battery charger defect
The battery charger is not work-
ing correctly
Contact XVIVO for service
Battery 1 not charging
Battery 2 not charging
Either battery 1 or 2 is not able
to hold a charge
Check wall power connection
Contact XVIVO for service
Bubble detected
An air bubble was detected in
the PA perfusion line
Check for bubbles
Reset bubble detector
Bubble sensor disconnected
Bubble sensor is not sending
signal to pump
Reconnect bubble sensor
Reset bubble detector
Countdown time elapsed
The countdown timer reached 0
Reset timer
Device defect (Ox$)
There is an internal conflict in
the pump system
Contact XVIVO for service
Fan 1 drive defect
Fan 2 drive defect
The fan is not operating proper-
ly
Contact XVIVO for service
Fan 1 housing defect
Fan 2 housing defect
The fan housing is defective
causing excessive resistance
Contact XVIVO for service
Fan AC supply defect
AC power to pump fans is not
working
Check wall power connection
Contact XVIVO for service
Flow sensor defect
The flow sensor is not sending
appropriate data to computer
Check flow sensor connection
Reconnect flow sensor
Re
-
zero flow sensor
Flow sensor disconnected
The flow sensor is working but is
not connected to the perfusion
tubing
Reconnect flow sensor
Re
-
zero flow sensor
Flow sensor offset too high
The flow sensor is not posi-
tioned properly in the perfusion
set
Reconnect flow sensor
Re
-
zero flow sensor
Level limit reached
The fluid level in the reservoir is
at the low point of the level
detector
Add STEEN Solution
TM
Level sensor disconnected
The level sensor is working but
is not connected to the disposa-
ble
Reconnect level sensor