XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
66
-
P/N: 9182
-
0189 Revision Q
Notice
Definition
Action needed
Ventilator Cdyn
Invalid Data
The dynamic compliance data
connection from the ventilator
is corrupt
If this value is invalid or out of range
use the following calculation Vte/PIP
-
PEEP.
Contact XVIVO to schedule service.
Ventilator Cdyn Out
of Range
The dynamic compliance data is
outside of normal limits
If this value is invalid or out of range
use the following calculation Vte/PIP
-
PEEP.
Contact XVIVO to schedule service
Ventilator Date/Time
Invalid Data
The date/time feature of the
ventilator is >5 minutes differ-
ent from the XPS software date/
time.
Adjust date/time of ventilator to
match XPS clock
Ventilator FiO2
Invalid Data
The oxygen percentage data
connection from the ventilator
is corrupt
Use ventilator readings for FiO2 for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator FiO2 Out
of Range
The FiO2 data is outside of nor-
mal limits
Use ventilator readings for FiO2 for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator Freq
Invalid Data
The frequency of breath data
connection from the ventilator
is corrupt
Use ventilator readings for Freq for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator Freq Out
of Range
The frequency of breath data is
outside of normal limits
Use ventilator readings for Freq for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator I:E Ratio
Invalid Data
The I:E Ratio data connection
from the ventilator is corrupt
Use ventilator readings for I:E ratio for
rest of procedure and ignore XPS soft-
ware.
Contact XVIVO to schedule service
Ventilator InspTime
Invalid Data
The inspiration time data con-
nection from the ventilator is
corrupt
Use ventilator readings for Inspiration
for rest of procedure and ignore XPS
software.
Contact XVIVO to schedule service
Ventilator InspTime
Out of Range
The inspiration data is outside of
normal limits
Use ventilator readings for Inspiration
for rest of procedure and ignore XPS
software.
Contact XVIVO to schedule service
5.2 XPS
TM
Software
Priming: Notice