XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
48
-
P/N: 9182
-
0189 Revision Q
4.3.4 Smart Sequencing
SMART SEQUENCING is an optional
feature that guides the user through
the EVLP protocol. It utilizes system
timers and specific parameters to
advance the user through each step
of the EVLP. A table of detailed steps
is found in the Appendix 1 Additional
Features, Section 8.9.
To initiate the SMART SEQUENC-
ING feature, donor parameters must
be entered, and SMART SEQUENC-
ING toggle switch must be ON.
Following PRIMING and RETRO-
GRADE MODES (sections 2.4 through
3.3.9), SMART SEQUENCING is initiat-
ed by starting Timer 1 (EVLP) as AN-
TEGRADE MODE starts.
As sequencing commences, the
current step will be indicated in the
SMART SEQUENCING window. Be-
tween each step a pop
-
up message
box will temporarily appear with a
audible sound to notify the user of a
change.
EVLP parameters to target for each
step of the protocol will be highlight-
ed on the setup page and displayed
as target values below specific pa-
rameters
Timer 2 (Ventilator) will automati-
cally start when ventilation is detect-
ed at the appropriate step and Timer
3 (Recruitment) will automatically
start when recruitments are detected
at the appropriate sequence.
NOTE 1: The XPS software program
can be run with SMART SEQUENCING
turned off. Parameters, timers and
trend graphs operate normally when
SMART SEQUENCING is disabled.
NOTE 2: If SMART SEQUENCING
needs to be aborted for any reason,
Timer 1 needs to be stopped (
),
followed by acceptance of an ABORT
pop
-
up message, then the SMART
SEQUENCING toggle switch (
) can
be turned OFF.