XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
72
-
P/N: 9182
-
0189 Revision Q
Notice
Definition
Action needed
Centrifugal Pump PA
Temperature Out of
Range
Temperature reading from PA
sensor is out of set range
Reset upper and/or lower temperature
notice ranges
Contact XVIVO to schedule service
Centrifugal Pump
Speed Invalid Data
Pump RPMs are not reading
properly
Make sure pump head/oxygenator is
seated correctly in pump housing
Replace oxygenator/pump head
Centrifugal Pump
Speed Out of Range
Pump RPMs are reading out of
set range
Make sure pump head/oxygenator is
seated correctly in pump housing
Reset upper and/or lower RPM notice
ranges
LA CO2 Below Limit
The CO2 reading for the LA is
reading below the set notice
limit
Check ventilator parameters, esp. rate,
FiO2
Reset the lower CO2 notice limit
Contact XVIVO to schedule service
LA CO2 PGM Out of
Range
The CO2 reading coming from
either the disposable sensor or
transmitter housing is out of
acceptable range
Check disposable to make sure sensor
is seated properly in housing.
Replace disposable sensor
Re
-
enter PGM calibration data
LA CO2 PGM Phase
Invalid Data
The CO2 reading does not
match calibration curve data
Re
-
enter PGM calibration data
Check disposable to make sure sensor
is seated properly in housing.
Replace disposable sensor, recalibrate
LA O2 PGM Out of
Range
The O2 reading coming from
either the disposable sensor or
transmitter housing is out of
acceptable range
Check disposable to make sure sensor
is seated properly in housing.
Replace disposable sensor
Re
-
enter PGM calibration data
LA pH Below Limit
The pH reading for the LA is
reading below the set notice
limit
Add fresh STEEN Solution
TM
to system,
remove similar volume of old STEEN
Solution
TM
Reset the lower CO2 notice limit
LA pH PGM Out of
Range
The pH reading coming from
either the disposable sensor or
transmitter housing is out of
acceptable range
Check disposable to make sure sensor
is seated properly in housing.
Replace disposable sensor
Re
-
enter PGM calibration data
UPS Hardware and/
or Communication
Problem
There is a hardware or loss of
communication problem with
the UPS to the XPS software.
Check UPS cable connection to
touchscreen monitor
Have the unit serviced
5.2 XPS
TM
Software
Priming: Notice