![XVIVO XPS Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/xvivo/xps/xps_instructions-for-use-manual_891588023.webp)
XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
23
-
P/N: 9182
-
0189 Revision Q
2.4.3 Flow/Bubble and
Level Sensors
FLOW/BUBBLE:
Place the flow/bubble sensor
around the PA (yellow
-
banded)
circuit tubing at a position close
to the duckbill check valve
without breaking the sterile
barrier as shown.
Close the cover with the
indicator pointing arrow toward
the lung and snap to lock.
LEVEL:
Affix disposable level sensor
pad to hard shell reservoir with
the arrow pointing to the right
and bottom edge of sensor lined
up near 50ml mark.
NOTE: The bottom edge of the
sensor indicates the point at
which the centrifugal pump will
stop should the fluid level drop
below the sensor.
Connect the level sensor to the
sensor pad.
NOTE: If fluid level should fall be-
low level sensor, pump will stop
operation until fluid level is re-
stored. See Section 2.4.2 for res-
ervoir filling instructions.