XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
17
-
P/N: 9182
-
0189 Revision Q
Connect STEEN Solution
TM
spike tubing to reservoir inlet
port
Locate the STEEN Solution
TM
drain tubing which comes off of
the Quadrox
-
iR oxygenator and is
not connected to anything else at
its terminal end and has an or-
ange
-
colored tape.
Feed the STEEN Solution
TM
drain tubing through second
(lower) pump head in the flow
direction shown by the arrow
(Right to Left)
Connect STEEN Solution
TM
drain tubing to drain bag at the
number
-
matched orange tape
stripe near luer port.
Locate the STEEN Solution
TM
recycle tubing which is a 24 inch
piece of tubing connected to the
drain bag with a red tape stripe.
Connect the open end of the
recycle tubing to a reservoir inlet
port and feed the STEEN Solu-
tion
TM
recycle tubing through the
third pump head in the direction
shown.
NOTE: Tubing must be inserted
from the RIGHT side (entry) to the
LEFT side (exit), centered across
peristaltic pump heads, carefully
fitting between the two black
tubing channel rails on both the
entry and exit ends of the pump
head or damage/ kinking to the
tubing may occur.
CAUTION: The adjustable black
settings on both sides of all 3
pump heads must be fixed in the
3rd click position down from the
top on each pump to properly
seat the tubing.
2.2.5 Connect Tubing
Lines to the XPS
TM
Cart
Disposable Set
-
up