Smart Sequencing
8.9 Smart Sequencing Table (cont.)
SEQUENCE
NUMBER
DESCRIP-
TION
TIME
RANGE
TEMP
TARGET
(C)
FLOW
TARGET
VENTILATION
PARAMETERS
Actions on
Automatic
Timers
ACTIVATES
NEXT SE-
QUENCE
POP
-
UP MESSAGE
AT START of PHASE
SMART SEQUENCE
MESSAGE
13
2nd Hour
Perfusion
Mode
62 min
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer at 1 Hour
48 Minutes
(NONE)
13: 2nd
-
HOUR
PERFUSION MODE
14
2nd Breath
Hold
108
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer at 1 Hour
48 Minutes: 30
Seconds
START 15
-
second
BREATH HOLD ON
VENTILATOR
14: 15
-
second
BREATH HOLD
15
2nd Recruit-
ment Notice
108.5
minutes
37
100% of
Max
RECRUITMENT
SETTINGS
(NONE)
PO2>80%
START 2nd RE-
CRUITMENT / O2
CHALLENGE
Vt = <Recruitment
VT>
Rate = 10
Oxygen = 100%
15: START 2nd
RECRUITMENT / O2
CHALLENGE
16
2nd Recruit-
ment Active
~109
minutes
37
100% of
Max
RECRUITMENT
SETTINGS
Timer 3
(Recruitment)
starts count-
down from 10
min
Timer 3
(Recruitment)
expired
(NONE)
16: 2nd RECRUIT-
MENT UNDERWAY
17
2nd
Recruitment
Ended
Notice
~119
minutes
37
100% of
Max
NORMAL
SETTINGS
Timer 3
(Recruitment)
at 0 min
PO2 < 40%
END RECRUITMENT
Vt = <Normal VT>
Rate = 7
Oxygen = 21%
17: END RECRUIT-
MENT
18
2nd Post
Recruitment
Perfusion
~120
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 2
Hours 48
Minutes
(NONE)
18: 3rd
-
HOUR
PERFUSION MODE
19
3rd Breath
Hold
168
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer at 2
Hours 48
Minutes: 30
Seconds
START 15
-
second
BREATH HOLD ON
VENTILATOR
19: 15
-
second
BREATH HOLD
20
3rd
Recruitment
Notice
168
minutes; 30
seconds
37
100% of
Max
RECRUITMENT
SETTINGS
(NONE)
PO2>80%
START 3rd RECRUIT-
MENT / O2 CHAL-
LENGE
Vt = <Recruitment
VT>
Rate = 10
Oxygen = 100%
20: START 3rd
RECRUITMENT / O2
CHALLENGE
21
3rd Recruit-
ment Active
~169
minutes
37
100% of
Max
RECRUITMENT
SETTINGS
Timer 3
(Recruitment)
starts count-
down from 10
min
Timer 3
(Recruitment)
expired
(NONE)
21: 3rd RECRUIT-
MENT UNDERWAY
22
3rd
Recruitment
Ended
Notice
~179
minutes
37
100% of
Max
NORMAL
SETTINGS
Timer 3
(Recruitment)
at 0 min
PO2 < 40%
END RECRUITMENT
Vt = <Normal VT>
Rate = 7
Oxygen = 21%
22: END RECRUIT-
MENT
23
3rd Post
Recruitment
Perfusion
~180
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 3
Hours 48
Minutes
(NONE)
23: 4th
-
hour PER-
FUSION MODE
24
4th Breath
Hold
228
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 3
Hours 48
Minutes: 30
Seconds
START 15
-
second
BREATH HOLD ON
VENTILATOR
24: 15
-
second
BREATH HOLD
25
4th
Recruitment
Notice
228
minutes; 30
seconds
37
100% of
Max
RECRUITMENT
SETTINGS
(NONE)
PO2>80%
START 4th RECRUIT-
MENT / O2 CHAL-
LENGE
Vt = <Recruitment
VT>
Rate = 10
Oxygen = 100%
25: START 4th
RECRUITMENT / O2
CHALLENGE
26
4th
Recruitment
Active
~229
minutes
37
100% of
Max
RECRUITMENT
SETTINGS
Timer 3
(Recruitment)
starts count-
down from 10
min
Timer 3
(Recruitment)
expired
(NONE)
26: 4th RECRUIT-
MENT UNDERWAY
27
4th
Recruitment
Ended
Notice
~239
minutes
37
100% of
Max
NORMAL
SETTINGS
Timer 3
(Recruitment)
at 0 min
PO2 < 40%
END RECRUITMENT
Vt = <Normal VT>
Rate = 7
Oxygen = 21%
27: END RECRUIT-
MENT
28
4th Post
Recruitment
Perfusion
~240
minutes
37
100% of
Max
NORMAL
SETTINGS
(NONE)
N/A
(NONE)
28: EXTENDED EVLP