XVIVO Perfusion System (XPS
TM
) Instructions for Use
-
102
-
P/N: 9182
-
0189 Revision Q
Smart Sequencing
8.9 Smart Sequencing Table
SEQUENCE
NUMBER
DESCRIP-
TION
TIME
RANGE
TEMP
TARGET
(C)
FLOW
TARGET
VENTILATION
PARAMETERS
ACTIONS ON
AUTOMATIC
TIMERS
ACTIVATES
NEXT SE-
QUENCE
POP
-
UP MESSAGE
AT START of PHASE
SMART SEQUENCE
MESSAGE
1
PRIMING
MODE
0 min
NONE
NONE
OFF
(NONE)
Timer 1 (EVLP)
Starts
N/A
1. PRIMING MODE.
Start EVLP Timer
when Antegrade
Perfusion starts.
2
Heating
Phase
0
-
10 min
22
10% of
Max
OFF
(NONE)
Timer 1 at 10
minutes
(NONE)
2. HEATING PHASE:
Set H/C to 23°C.
FLOW TARGET
<Target Flow (Start)
> LPM
3
Heating
Phase
10
-
20 min
31
20% of
Max
OFF
(NONE)
Timer 1 at 20
minutes
10 MINUTE NOTIFI-
CATION:
Set H/C to 32°C
Target Flow of:
<ActualFlow_10_Mi
ns>
3. HEATING PHASE:
Set H/C to 32°C.
FLOW TARGET
<Target Flow
(10min)> LPM
4
Heating
Phase
20
-
30 min
32
30% of
Max
NORMAL
SETTINGS
(NONE)
LA Temp at 32
20
-
MINUTE NOTIFI-
CATION:
Set H/C to 38°C
Target Flow of:
<ActualFlow_20_Mi
ns>
4. HEATING PHASE:
Set H/C to 38°C.
FLOW TARGET
<Target Flow (20
mins)> LPM
5
Start
Ventilation
Notice
20
-
30 min
32
30% of
Max
NORMAL
SETTINGS
(NONE)
Ventilation
Detected
"VENTILATION AND
DEOXYGENATION
GAS READY TO
START
Set H/C to 38°C
Target Flow of:
<ActualFlow_20_Mi
ns>
Vt = <NORMAL
VT>
Rate = 7
5. START VENTILA-
TION AND DEOX
GAS
Set H/C to 38°C
FLOW TARGET
<Target Flow (20
Min) > LPM
6
Ventilation
Detected
20
-
30 min
32
30% of
Max
NORMAL
SETTINGS
Timer 2
(Ventilation)
Starts count
-
up
Timer 1 at 30
minutes
(NONE)
6. HEATING PHASE:
VENTILATION ON
Set H/C to 38°C
FLOW TARGET
<Target Flow (20
Min)> LPM
7
Heating
Phase
30
-
40 min
37
50% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 40
minutes
30
-
MINUTE NOTIFI-
CATION:
Target Temp 37°C
Target Flow of:
<ActualFlow_30_Mi
ns>
7. HEATING PHASE:
VENTILATION ON
TEMP TARGET 37°C.
FLOW TARGET
<Target Flow (30
Min)> LPM
8
Heating
Phase
40
-
50 min
37
80% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 50
minutes
40
-
MINUTE NOTIFI-
CATION:
Target Flow of:
<ActualFlow_40_Mi
ns>
8. HEATING PHASE:
VENTILATION ON
FLOW TARGET
<Flow Target (40
Min)> LPM
9
Full Flow
then Breath
Hold
50 min
37
100% of
Max
NORMAL
SETTINGS
(NONE)
Timer 1 at 50.5
minutes
50
-
MINUTE NOTIFI-
CATION:
Target Flow of:
<ActualFlow_50_Mi
ns>
START 15
-
second
BREATH HOLD ON
VENTILATOR
9. HEATING PHASE:
15 SECOND BREATH
HOLD
10
First Re-
cruitment
Notice
50.5 min
37
100% of
Max
RECRUITMENT
SETTINGS
(NONE)
PO2>80%
START 1st RECRUIT-
MENT / O2 CHAL-
LENGE
Vt = <Recruitment
VT>
Rate = 10
Oxygen = 100%
10. START 1st
RECRUITMENT / O2
CHALLENGE
11
First
Recruitment
Active
~51
-
61min
37
100% of
Max
RECRUITMENT
SETTINGS
Timer 3
(Recruitment)
starts count-
down from 10
min
Recruitment
Timer expired
(NONE)
11: 1st RECRUIT-
MENT UNDERWAY
12
First
Recruitment
Ended
Notice
~61 min
37
100% of
Max
NORMAL
SETTINGS
Timer 3
(Recruitment)
at 0 min
PO2 < 40%
END RECRUITMENT
Vt = <Normal VT>
Rate = 7
Oxygen = 21%
12: END RECRUIT-
MENT