) Instructions for Use
-
70
-
P/N: 9182
-
0189 Revision Q
Notice
Description
Action needed
PA CO2 PGM phase
data corruption
Data from PA CO2 sensor is
Inconsistent (multiple data points)
Check position of PGM disposable
sensor and re
-
seat if necessary
PA pH PGM phase
data corruption
Data from PA pH sensor is Incon-
sistent (multiple data points)
Check position of PGM disposable
sensor and re
-
seat if necessary
Pulmonary artery
CO2 monitor com-
munication lost
Data connection between XPS and
PA PGM sensor disrupted
Check position of PGM disposable
sensor and re
-
seat if necessary
Check if sensor spot is missing
Replace disposable PGM sensor
Pulmonary artery O2
monitor communica-
tion lost
Data connection between XPS and
PA PGM sensor disrupted
Check position of PGM disposable
sensor and re
-
seat if necessary
Check if sensor spot is missing
Replace disposable PGM sensor
Pulmonary artery pH
monitor communica-
tion lost
Data connection between XPS and
PA PGM sensor disrupted
Check position of PGM disposable
sensor and re
-
seat if necessary
Check if sensor spot is missing
Replace disposable PGM sensor
Recycle STEEN pump
communication lost
Connection to peristaltic
‘
Recycle
STEEN
’
pump disrupted
Drain STEEN Solution
TM
by hand
removing from dome drain port
with large (100cc +) syringe con-
nected to luer port
Reboot software if necessary
Contact XVIVO for service
Ventilator communi-
cation lost
Ventilator was communicating
with XPS but then communication
was lost
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
Ventilator date/time
corruption
The date/time on the ventilator is
5 minutes > or < than XPS time
Reset ventilator date/time
Ventilator freq data
corruption
Data between XPS and ventilator
disrupted
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
Ventilator I:E ratio
data corruption
Data between XPS and ventilator
disrupted
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
Ventilator Peak data
corruption
Data between XPS and ventilator
disrupted
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
Ventilator PEEP data
corruption
Data between XPS and ventilator
disrupted
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
Ventilator VTE data
corruption
Data between XPS and ventilator
disrupted
Confirm ventilator power is on
and standby mode is
‘
off
’
Use ventilator controls if working
Reboot software if possible
Contact XVIVO for service
5.2 XPS
TM
Software
XVIVO Perfusion Cart Software
Priming: Notice