34
World Precision Instruments
Command
Syntax
Notes
Display syringe type
?S
Response: Type 8, 100.0uL, 60.0
OK
Display direction type
?D
Response: Direction: Infuse
(Direction: Withdraw)
OK
Display rate units
?U
Response: Rate Units: nL/min
(Rate Units: nL/sec)
OK
Display run mode
?G
Response: Motor State: Stopped
(Motor State: Running, Motor State: Paused)
OK
Pause script execution
A####
#### is length of pause in seconds/100
Response: PAUSING (while in pause)
OK (when finished pause)
Beep
F####
#### is length of beep in seconds/100
Response: BEEP (while beeping)
OK (when finished beeping)
Kill command
Z
Ends current command and cancels any
pending commands
Blocking start delivery
*G
Same as G command, but it stops accepting
any new commands until the commanded
delivery is finishd at which point it responds
with OK. This is useful when host computer
wants to ensure delivery is finished before
commanding another action.